(337b) Fabrication of Enzymatic Electrochemical Biosensors for Real-Time Nicotine Detection

Authors 

Chen, M. - Presenter, Boston University
Kuzmanovic, U. - Presenter, Boston University
Sensharma, P., Boston University
Gupta, A., Boston University
Sankar, K., Boston University
Tararina, M., Boston University
Galagan, J., Boston University
Allen, K., Boston University
Grinstaff, M., Boston University
Biosensors provide an output signal based upon a successful recognition of an analyte. Key to this process in oxidase-based biosensor is analyte recognition by an enzyme and subsequent signal transduction. As such, the sensor design process must successfully address a series of requirements associated with the interface between the enzyme and an electrode. Here we present the development, assessment, and optimization of a nicotine sensor based on a nicotine catabolizing redox enzyme and Prussian Blue coated screen-printed electrodes to afford a chronoamperometric response proportional to nicotine. An optimized polymeric matrix immobilization method maintains enzymatic activity on the mediator-coated screen-printed electrode and enables quantitative detection of nicotine. The resulting biosensor detects nicotine in urine with high sensitivity with a limit of detection of 34 μM over the range of 0-200 μM.